External PD Calibrator

Home >> Technical >> External PD Calibrator

External PD Calibrator Type HVS 998/0011

The calibration of a PD test circuit is a necessary step in the test procedure. The system is fed with a calibration pulse of a known value of apparent charge.

  • Battery powered
  • Synchronization with mains frequency
  • Adjustable apparent charge

Design Features

The calibrator Type HVS 998/0011 feeds a short duration current pulse of a known magnitude through the terminals of either the de-energized test objet [direct calibrationJ or the coupling device (indirect calibration) The PD detector will be adjusted to the value of the apparent charge given by the calibrator . Synchronization with the line Frequency is supported through optical coupling. The calibrator complies with IEC 270.

Order specification

HVS 998/0011 external PD calibrator

Technical Specifications Calibrator Type HVS 998/0011

Calibrator
Minimum
Maximum
Unit
Pd calibration pulse
10
10000
pC
Range adjustment
10
100
%
Repetition
90
900
Hz
Phase angle
0
Deg
Pulse rise time
<100
nS
Pulse duration
>20
μS
Pulse trailing slope time
500
μS


Environmental
Minimum
Maximum
Unit
Battery voltage
9
V
Temperature range
5
40
°C
Relative humidity
<80
%


Mechnical
Minimum
Maximum
Unit
Dimensions
67x110x42
mm
Instrument weight
0.2
kg